(Master – to be added to the documentation)
EN
4 - 4 Infusomat® compact
plus
P 1.0
For internal use, only
4 Check after Repair
Checklist for Check after Repair
Checklist for Check after Repair – After Repair or Calibration Owner
Unit: Infusomat® compact
plus
P
Manufacturer: B. Braun Melsungen AG
Year of Procurement
Observe the Service Manual and the Instructions for Use. All measured values are
to be documented. Accessories used should be included in testing. Only use calibrated
measuring equipment.
Software Version
Article No. (REF) Serial No. (SN) Inventory No. (of the Owner)
YES N/A
T1 Visual Inspection ❒❒
Infusomat® compact
plus
P:
Cleanliness, completeness, damage,
and defects affecting safety, damage
and readability of the labels.
Particularly:
Housing
Screw covers
Pump door
SHK ICPP
Air sensor (free from grease or
ultrasonic gel, free from cracks)
Pressure sensor, downstream
(membrane)
Pump membrane
Pressure sensor, upstream
(membrane)
Second clamp
Keypad/display
Accessory connector
IR window
Mains power supply connection
Accessories
Cleanliness, completeness, damage,
and faults affecting safety, damage,
and readability of the labels
Check the unit and the accessories for
compatibility
Mains power supply cable
YES N/A
T2 Electrical Safety ❒❒
Protective conductor resistance of
mains connecting line
( 0.1 Ω) ______ Ω
Measure
mains voltage _____ V~
Equipment leakage current
(alternative method for devices of
protection class II)
(≤ 10 μA) _____ μA
Measure between mains input and
pin 3 of accessory connector. Use the
Measuring Adapter Accessory
Connector CP.
YES N/A
T3 Mechanical Inspection ❒❒
Locking with second unit
Pole clamp
YES N/A
T4 Operational Test ❒❒
Connect unit to mains power supply
Status display
Switch on unit with mains power supply
Self-test
Information on display
Audible alarm
Visual alarm
YES N/A
T4 Operational Test (cont.) ❒❒
Switch on unit without mains power supply
Self-test
Battery test
Opening and closing mechanism of the
operating unit
Infusion
Buttons on the operating unit
Trigger bolus at the device
Staff call (if available)
YES N/A
T5 Pressure cut-off, electronic ❒❒
Delivery rate: 250 ml/h.
Upstream
Alarm
Downstream
Pressure stage 5
(0.1 … 0.7 bar) _____ bar
(or 1.5 … 10 psi) _____ psi
Pressure stage 9
(0.7 … 1.3 bar) _____ bar
(or 10 … 19 psi) _____ psi
YES N/A
T6 Pressure limitation, mechanical ❒❒
Delivery rate: 250 ml/h.
P
max
(2.4 … 4.0 bar) _____ bar
(or 35 … 58 psi) _____ psi
(Sheet 1 of 2)