2 Unit Diagnosis / Adjustments
2 - 6 Infusomat® compact
plus
P 1.0
EN
For internal use, only
THE MOST IMPORTANT FAULT EVENTS The following list specifies the most important known errors and
how they are rectified.
76 Invalid voltage motor Hardware defect
200-202 Air-sensor supervision Hardware defect
206 Light barrier supervision Hardware defect
207 Opening / closing of the clamp (SHK,FFC)
is not detected by the photocell
Hardware defect
209 Air-sensor calibration data invalid Hardware defect
210 Air-sensor flash data invalid Hardware defect
211 Air-sensor hardware version invalid
213 Air-sensor in service mode Software failure.
If persisting: hardware defect
214 Pressure-sensor values out of range Hardware defect
215 Sensor PCB hardware version invalid
216 Temperature supervision Hardware defect
Code Definition Possible cause Fault rectification
Table 2 - 3 Error codes (Cont.)
Error Possible cause Fault rectification
1. The battery is discharging too
fast.
The device has not been used for a longer
period of time. The battery has not been
discharged and recharged on a regular basis.
Discharge and recharge battery several
times.
Replace battery module.
Table 2 - 4 Most important fault events
NOTICE
The unit must be checked after every repair or service.
(see "Checks after Repair" ➡ p. 3 - 32)