Infusomat® compact
plus
P 1.0 6 - 1
For internal use, only
6 Technical Safety Check (TSC)
Checklist for Technical Safety Check (TSC)
Checklist for Technical Safety Check – Every 24 Months Owner
Unit: Infusomat® compact
plus
P
Manufacturer: B. Braun Melsungen AG
Year of Procurement
Observe the Service Manual and the Instructions for Use. All measured values are
to be documented. Accessories used should be included in testing. Only use calibrated
measuring equipment.
Software Version
Article No. (REF) Serial No. (SN) Inventory No. (of the Owner)
Visual Inspection Electrical Safety
According to IEC 62353
Functional Inspection
❒ Infusomat® compact
plus
P:
Cleanliness, completeness, damage,
and defects affecting safety, damage
and readability of the labels.
Particularly:
❒ Housing
❒ Screw covers
❒ Pump door
❒ SHK ICPP
❒ Air sensor (free from grease or
ultrasonic gel, free from cracks)
❒ Pressure sensor, downstream
(membrane)
❒ Pump membrane
❒ Pressure sensor, upstream (mem-
brane)
❒ Second clamp
❒ Keypad/display
❒ Accessory connector
❒ IR window
❒ Mains power supply connection
❒ Accessories
Cleanliness, completeness, damage,
and faults affecting safety, damage,
and readability of the labels
❒ Check the unit and the accessories for
compatibility
❒ Mains power supply cable
❒ Protective conductor resistance of
mains connecting line
(≤ 0.1 Ω) ______ Ω
❒ Measure
mains voltage _____ V~
❒ Equipment leakage current
(alternative method for devices of
protection class II)
(≤ 10 μA) _____ μA
Measure between mains input and
pin 3 of accessory connector. Use the
Measuring Adapter Accessory Con-
nector CP.
Mechanical Inspection:
❒ Locking with second unit
❒ Pole clamp
Operational Test:
Connect unit to mains power supply
❒ Status display
Switch on unit with mains power supply
❒ Self-test
❒ Information on display
❒ Audible alarm
❒ Visual alarm
Switch on unit without mains power supply
❒ Self-test
❒ Battery test
❒
Opening and closing mechanism of the
operating unit
❒ Infusion
❒ Buttons on the operating unit
❒ Trigger bolus at the device
❒ Staff call (if available)