2 Unit Diagnosis / Adjustments
2 - 2 Infusomat® compact
plus
P 1.0
EN
For internal use, only
STANDARD DEVICE ALARMS Appearance
Header with short information (e.g., “Device alarm“) and 4-digit
error code (e.g., „0325“). Footer with on/off-icon and short text
(e.g., “Turn Pump off“).
Error codes of the function microprocessor (FuP)
Error codes of the control microprocessor (KuP)
NOTICE
In case a device alarm appears, please switch the device off and
on. Further action only needs to be taken if the alarm persists.
Code Definition Possible cause Fault rectification
319 Validation of the Displayed Disposable
failed
Software failure
325 Key panel activated Key panel failure Change upper part of housing
326 Key panel not confirmed by KuP Hardware defect Change upper part of housing
350 Pressure values delayed Software failure
358 Unexpected error in the timer
management of the pump
Software failure
359 Deviation during periodic "Readback" of
KuP data
Software error, e.g due to corrupted data in
RAM
Table 2 - 1 Error codes of the FuP
Code Definition Possible cause Fault rectification
704 Disposable table damaged Software failure
705 Disposable data not plausible Software failure
706 Disposable table damaged Software failure
707-709 Plausibility of sensor data concerning
status of the pump invalid
Software failure
719 Change of pressure value, user
confirmation not detected by KuP
Software failure
735 Self check not ready on start of infusion Software failure
740 Battery voltage inconsistent Battery defect Change battery
Table 2 - 2 Error codes of the KuP