4 Check after Repair
Infusomat® compact
plus
P 1.0 4 - 3
EN
For internal use, only
8. Select "RUN IN" and confirm with "OK" (Fig. 4 - 3).
Line calibration is started.
Fig. 4 - 3
NOTICE
This service function is automatically deactivated again after 5
minutes. The RUN IN has to be selected within these 5 minutes.
After selection of RUN IN, the RUN IN line remains active until
the next line change.