4 Check after Repair
4 - 14 Infusomat® compact
plus
P 1.0
EN
For internal use, only
5. Enter a delivery rate of 250 ml/h and a volume of 250 ml.
6. Start infusion.
7. Generate an air bubble of approx. 0.4 ml (approx. 56 mm length
of line with air) in the supply line (injection port) to the pump.
An alarm is triggered when the air bubble is detected.
8. Insert an air-filled Intrafix Primeline into the device and close
the operating unit, or disconnect the line at the container and
“empty“, so that there is no longer any water in the line.
9. Display sensor information via "Sensor info" menu item.
10. Wait at least 120 seconds.
11. Read off the "Air Voltage" value and document.
T11 Temperature sensor
1. Display the air sensor information (see "T10 Air sensor"
➡ p. 4 - 13).
2. Read off the values of the FuP and KuP temperature sensors on
the LC display and compare the difference with the
specifications in the checklist.
3. Calculate the temperature differences according to the
checklist and document.
T12 Staff Call Test
1. Connect the staff call cable compact
plus
to the accessory
connector of the pump.
2. Insert the Intrafix Primeline and start the pump with an
arbitrary delivery rate.
3. Open the pump door.
4. An alarm goes off and the red LED of the staff call cable lights
up.
Fixing the Seal
1. After each repair do not forget to re-seal the device with your
individual seal.
NOTICE
If the water or air value is not reached immediately, these values
must be read again after 120 seconds.
NOTICE
If the water or air value is not reached immediately, these values
must be read again after 120 seconds.