7 Procedural Instructions on the TSC
Infusomat® compact
plus
P 1.0 7 - 9
EN
For internal use, only
5. Enter a delivery rate of 250 ml/h and a volume of 250 ml.
6. Start infusion.
7. Generate an air bubble of approx. 0.4 ml (approx. 56 mm length
of line with air) in the supply line (injection port) to the pump.
An alarm is triggered when the air bubble is detected.
8. Insert an air-filled Intrafix Primeline into the device and close
the operating unit, or disconnect the line at the container and
“empty“, so that there is no longer any water in the line.
9. Display sensor information via "Sensor info" menu item.
10. Wait at least 120 seconds.
11. Read off the "Air Voltage" value and document.
Temperature sensor
1. Display the air sensor information (see "Air sensor" ➡ p. 7 - 8).
2. Read off the values of the FuP and KuP temperature sensors on
the LC display and compare the difference with the
specifications in the checklist.
3. Calculate the temperature differences according to the
checklist and document.
NOTICE
If the water or air value is not reached immediately, these values
must be read again after 120 seconds.
NOTICE
If the water or air value is not reached immediately, these values
must be read again after 120 seconds.