UF V (E3) 03/2019 Page 102/116
Chamber size 500 700
Electrical Data UF V (230 V)
IP system of protection acc. to EN 60529 IP 20 20
Nominal voltage (±10%) at 50 Hz power frequency V 230 230
Current type 1N~ 1N~
Nominal power kW 1.6 1.6
Nominal current A 7.0 7.0
IEC connector plug and cable mm / inch 2000 / 78.7 2000 / 78.7
Power plug Shock-proof plug
Installation category acc. to IEC 61010-1 II II
Pollution degree acc. to IEC 61010-1 2 2
Over-current release category C, 2 poles A 10 10
Electrical Data UF V-UL (115 V)
IP system of protection acc. to EN 60529 IP 20 20
Nominal voltage (±10%) at 60 Hz power frequency V 115 115
Current type 1N~ 1N~
Nominal power kW 1.4 1.4
Nominal current A 11.7 11.7
Power cable (IEC connector plug) mm / inch 2000 / 78.7 2000 / 78.7
Power plug NEMA 5-20 P 5-20 P
Installation category acc. to IEC 61010-1 II II
Pollution degree acc. to IEC 61010-1 2 2
Over-current release category C, 2 poles A 13 13
Electrical Data UF V UL (208 - 240 V)
IP system of protection acc. to EN 60529 IP 20 20
Nominal voltage (±10%) at 60 Hz power frequency V 208 - 240 208 - 240
Current type 2~ 2~
Nominal power kW 1.6 1.6
Nominal current A 7.7 7.7
IEC connector plug and cable mm / inch 2000 / 78.7 2000 / 78.7
Power plug NEMA 6-20P 6-20P
Installation category acc. to IEC 61010-1 II II
Pollution degree acc. to IEC 61010-1 2 2
Over-current release category C, 2 poles A 10 10
Environment-specific data UF V (230 V)
Noise level (mean value) dB (A) 47 47
Energy consumption at -80 °C/ -112 °F, with an
ambient temperature of +25 °C / 77°F; ± 10%
kWh/day 7,9 8,1
Heat dissipation at set-point -80 °C / -112 °F Wh/h 330 340
Filling weight of refrigerant R290 (propane)
(1
st
stage cooling, GWP 3)
kg 0.15 0.15
Filling weight of refrigerant R170 (ethane)
(2
nd
stage cooling, GWP 6)
kg 0.15 0.15
UF V with water cooling: Filling weight of refrigerant
R290 (propane) (1
st
stage cooling, GWP 3)
kg 0.15 0.15
UF V with water cooling: Filling weight of refrigerant
R170 (ethane) (2
nd
stage cooling, GWP 6)
kg 0.15 0.15