5.1 Menu structure of the controller and access levels
Starting from Normal display, navigate between the menus with the arrow buttons.
With the OK button you enter the setting of further subordinate menu functions.
With the Back button you go back to the previous function and finally back to Normal display.
The available functions depend on the current dependent on the current authorization “User”, “Admin” or
“Service”, for which the entry of a password may be required, depending on the setting.
You can set passwords for different access levels:
User: The password enables access to the standard operating functions. Factory setting is 00 00 (no
password assigned).
Admin: The password enables access to advanced controller functions and settings. Factory setting is
00 01.
Service: The password enables access to all controller functions (for BINDER Service only).
As soon as a password has been assigned, access to the respective functions is blocked and only availa-
ble after entering the correct password.
Menu Required access level Functions
Setpoints
“User
Tem peratu re set -point setting
CO
2
emergency cooling set-point setting (option
Setting the safety controller
Activating/deactivating the CO
2
emergency cooling (op-
tion) and the service setpoint, emergency cooling test (op-
tion)
Chamber info
Any user
Configuration display (setup information, controller hard-
ware and software, analog inputs)
Display of battery condition (“battery-buffered alarm sys-
temoption)
Display of interface configuration (e.g. MAC address, IP
address)
Settings
“Admin”
General controller settings (date, time, menu language,
temperature unit, display brightness)
Network settings
Setting the data logger storage interval
Setting the tolerance ranger limits and delay times for door
open and tolerance range alarm, setting the service set-
point
Access code assignment (“door access systemoption)
Password changing for User and Admin
Service
“Service”
Configuration settings (only for BINDER Service)
Password changing for User and Admin
USB
Export: Any user
Import: Admin
Export of configuration, logger, and service data
Import of configuration data