Binder UF V - Messages on the Battery Management System (Chamber with Option Battery-Buffered Alarm System)

Binder UF V
117 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
UF V (E3) 03/2019 Page 59/116
14.4.5 Messages on the battery management system (chamber with option battery-
buffered alarm system”)
The alarms and information messages occur immediately.
Battery operation
Alarm message Battery operation”, “collective alarm” icon
Audible alarm (buzzer)
Switching the zero-voltage relay alarm output
The message occurs on battery power (during power failure) after pressing the OK button (see chap.
10.3)
There is sufficient battery voltage.
Battery missing, completely empty (< 3V) or defective.
Alarm message Battery defective“, collective alarm icon
Audible alarm (buzzer)
Switching the zero-voltage relay alarm output
The message occurs on line power when no battery has been recognized.
Actions:
With defective battery: replace battery.
Reset alarm message. It will not reset automatically.
Low battery voltage
Information messageLow battery voltage, Information icon
The message occurs on battery power and on line power.
Battery voltage is very low (< 12 V). If this message lasts for more than 2 hours, the battery can be defec-
tive.
Actions:
Check the power supply.
Check the charging voltage in the controller menu (chap. 19.2)
With defective battery: replace battery. Contact BINDER Service.
Reset alarm message. It will not reset automatically.