UF V (E3) 03/2019 Page 55/116
Condition Alarm message
Moment of Alarm message and
switching the zero-voltage relay
alarm output
Safety controller temperature sen-
sor defective
Safety control sensor immediately
Inner temperature sensor and safe-
ty controller temperature sensor
defective. Refrigeration is turned on
permanently.
Temperature display
shows “- - - - “ or
“<-<-< “ or >->->”
Messages alternating:
Inner temp. sensor and
Safety control sensor
immediately
Failure of Pt100 temperature sen-
sor on the condenser of the refrig-
erating machine. Refrigeration is
turned on permanently.
Condensate temp.
sensor
immediately
Failure of Pt100 temperature sen-
sor on the cascade of the refrigerat-
ing machine. Refrigeration is turned
on permanently.
Cascade temp. sensor immediately
Failure of Pt100 temperature sen-
sor for ambient temperature or air
intake of the 1
st
stage cooling (re-
frigerating machine). Refrigeration
is turned on permanently.
Ambient temp. sensor immediately
Chamber with options battery-
buffered alarm systemand CO
2
emergency cooling: Battery opera-
tion (during power failure)
Battery operation immediately
Chamber with optional battery-
buffered alarm system”: No battery
recognized. Cause: Battery missing
or defective. Message on line pow-
er.
Battery defective immediately
Chamber with optional CO
2
emer-
gency cooling: activated CO
2
emer-
gency cooling automatically turned
on due to power failure and is run-
ning.
“CO2 emergency cooling” immediately
Chamber with optional CO
2
emer-
gency cooling: Insufficient CO
2
supply of the activated CO
2
emer-
gency cooling system
CO2 pressure immediately
Power failure (chamber without
battery or with message Battery
defective”)
-- immediately (alarm output only)
Press the OK button to confirm the alarm.
Confirmation while the state of alarm persists: Only the buzzer is muted. The visual alarm message
continues to be displayed until the alarm condition is removed. Then it is reset automatically.
Confirmation after the alarm has ended: The buzzer and the visual alarm message are rest together.