EasyManua.ls Logo

Biotage Selekt - Data Administration; Administrate Column Types; Administrate Solvents

Biotage Selekt
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...

Data Administration
Data Administration
Note: Only users with system owner privilege can administrate
column types, solvents, rack types, and user accounts.
Administrate Column Types
The software comes with a preconfigured list of column types
and their settings. To add, edit, and delete user-defined column
types, press Menu, Data Administration and then Column
Administration (see Figure 46).
Figure 46. The Column Administration view.
Column Parameters
Warning
»
Always use Biotage Safety Valve (P/N 417115SP) when
processing columns or combinations (e.g. column and dry
load vessel) with a total CV larger than 0.8 L.
»
Never use columns or combinations with a total CV larger
than 3.1 L.
»
Never exceed the maximum pressure or flow rate for the column.
Note: The preconfigured Biotage column types cannot be edited
or deleted. We recommend that you disable Visible for any
column that will not be used.
»
Visible: If enabled, the column type is available for use.
»
Name: The name of the column type.
»
Column Volume: The column volume (CV) in mL.
»
Default Flow Rate: The default flow rate in mL/min.
»
Max Pressure: The safety pressure in bar or psi
(see page 27). If reached, the flow rate will be
reduced to keep the pressure at this level.
»
Equilibration Pressure: The maximum pressure during the
equilibration, in bar or psi (see page 27). If reached, the
flow rate will be reduced to keep the pressure at this level.
»
Equil./Equilibration Length: The equilibration length in CV.
»
Equilibration Flow Rate: The equilibration
flow rate in mL/min.
»
Air Flush Length: The air flush length in CV.
This value will be used as the default volume
for the air flush feature (see page 24).
»
Load Capacity at CV=2: The approximate
load capacity when delta CV is 2.
»
Chemistry Type: The type of chromatography the
column is used for, normal or reversed phase.
»
Part Number: The manufacturer’s part number.
Administrate Solvents
The software comes with a preconfigured list of solvents
and their settings. To add, edit, and delete user-defined
solvents, press Menu, Data Administration and then Solvent
Administration (see Figure 47).
Figure 47. The Solvent Administration view.
Solvent Parameters
Note: The preconfigured Biotage solvents cannot be edited or
deleted. We recommend that you disable Visible for any solvent
that will not be used.
»
Visible: If enabled, the solvent is available for use.
»
Name: The solvent name.

Table of Contents

Other manuals for Biotage Selekt

Related product manuals