EasyManua.ls Logo

Biotek MicroFlo select - Page 111

Biotek MicroFlo select
130 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Error Codes | 93
MicroFlo Select Microplate Dispenser
Error Name Code Brief Description
BOOT_POWERUP_CHECKSUM_FAIL 1001 Bootcode powerup checksum test failed
BOOT_ERROR_UNKNOWN 1002 Unknown error in bootcode
BOOT_PAGE_PROGRAM_ERROR 1003 Bootcode page program error
BOOT_BLOCK_SIZE_ERR 1004 Bootcode block size error (not 256)
BOOT_INVALID_SIGNATURE 1005 Invalid processor signature (not
1280,1281,2560,2561)
BOOT_MEMORY_EXCEEDED 1006 Bootcode memory exceeded
BOOT_INVALID_SLAVE_PORT 1007 Invalid slave port
BOOT_INVALID_SLAVE_RESPONSE 1008 Invalid response from slave
BOOT_INVALID_PROCESSOR 1009 Invalid processor detected
BOOT_DOWNLOAD_CHECKSUM_ERR 1010 Checksum error downloading basecode
PARAMETER_LIMIT_ERR 2400 Parameter limit exceeded
PROGRAM_LOCKED 4000 Program locked so operation denied
PROGRAM_NONERASABLE 4100 Program non erasable so delete denied
SC_NAK_ERROR 8100 Communications NAK
SC_TIMEOUT_ERROR 8101 Timeout while waiting for serial message
data
SC_COM_BUSY_ERROR 8102 Instrument busy and unable to process
message
SC_RX_BUF_OVERFLOW_ERROR 8103 Receive buffer overflow error
SC_CHECKSUM_ERROR 8104 Checksum error
SC_INVALID_SCRUCT_TYPE 8105 Invalid structure type in byMsgStructure
header field
SC_INVALID_DESTINATION 8106 Invalid destination in byMsgDestination
header field
SC_OBJECT_NOT_SUPPORTED 8107 Request object received not supported by
instrument
SC_MSG_BODY_SIZE_ERROR 8108 Message Body size exceeds max limit
SC_MAX_REQS_RUNNING_ERROR 8109 Max number of requests currently running
and cannot run the latest request
SC_NO_REQS_RUNNING_ERROR 810A No request running when response
request issued
SC_RESP_NOT_READY_ERROR 810C Response for outstanding request not
ready yet
SC_INST_NOT_IN_PC_COMM_MODE 810D Instrument not in a mode to receive a PC
message
SC_REQ_PARAMS_NOT_VALID 810E One or more request parameters are not
valid
TCB_NOT_AVAIL_ERR A000 Task control block not available
DEV_NOT_AVAIL_ERR A100 <device> not available
CODE_VERSION_ERR A200 Version strings for multiple uProcessors
do not match
POWER_SUPPLY_ERR A300 <test type> power supply level error
MALLOC_ERR A400 Malloc failed
DISPLAY_IN_USE A500 Multiple tasks attempted to use display
simultaneously

Table of Contents