8 | Introduction
BioTek Instruments, Inc.
Dispense Precision and Accuracy
Cassette Precision Accuracy
< 10%CV @ 1L per well ± 10% @ 1L per well
1µL
< 5%CV @ 2L per well* ± 5% @ 2L per well*
< 10%CV @ 1L per well ± 10% @ 1L per well
1µL 1536
< 5%CV @ 2L per well* ± 5% @ 2L per well*
< 5%CV @ 5L per well ± 4% @ 5L per well
5µL
< 2.5%CV @ 10L per well* ± 2% @ 10L per well*
< 4%CV @ 10L per well ± 4% @ 10L per well
10µL
< 2%CV @ 20L per well* ± 2% @ 20L per well*
* These specifications are for these dispense volumes and higher.