Recommended Qualification Schedule | 71
MicroFlo Select Microplate Dispenser
Performance Qualification (PQ) confirms that the dispenser consistently meets the
requirements of the tests performed at your laboratory.
The recommended PQ procedure consists of performing the System Self-Check,
the Checksum Test, and the Dispense Precision and Accuracy Test.
Your facility’s operating policies may also require that you routinely perform
an actual assay, to confirm that the dispenser will consistently give adequate
results for the assays to be run with it.
These tests should be performed routinely; the recommended interval is
monthly. This frequency may be adjusted depending on the trends observed
over time.
The successful completion of the PQ procedure confirms that the dispenser is
performing consistently under normal operating conditions.
Recommended Qualification Schedule
IQ OQ PQ
Task/Test
Initially
Initially &
Annually
Monthly
Installation and Setup
System Self-Check
Checksum Test
Dispense Precision and Accuracy Test
Run Assay