EasyManua.ls Logo

Biotek MicroFlo select - Specifications

Biotek MicroFlo select
130 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specifications | 7
MicroFlo Select Operator's Manual
Specifications
Labware
96-, 384-, 1536-well
Microplates
6-, 12-, 24-, 48-well
Standard, low profile, or deep-well depth with
outside dimensions of 5.030” 0.010” x
3.365” 0.010” (127.76 mm 0.25 mm x
85.47 mm 0.25 mm)
Microstrips
1 x 8, 1 x 12 Standard
Microwells
Flat, round, “V” bottom, PCR, or half-area
12 mm X 75 mm
Test tubes
13 mm X 100 mm
Test tube rack accessory available from BioTek
is required (PN 7170520)
Hardware & Environmental
Dead Volume
Cassette
Types
Dispense range Cassette Life
24” tubing 30” tubing
1 µL
1 – 50µL in 1µL
increments
1000 384-well plates
@ 5µL/well
.96 mL 1.2 mL
1 µL 1536
1 – 50µL in 1µL
increments
250 1536-well plates
@ 5µL/well
.96 mL 1.2 mL
5 µL
5 – 2500µL in 5µL
increments
1000 96-well plates
@ 50µL/well
3.4 mL 4.2 mL
10 µL
10µL – 10mL in 10µL
increments
1000 96-well plates
@ 100µL/well
5.9 mL 7.4 mL
Peristaltic pump:
Positive displacement peristaltic pump with 4 rollers that stretch
the 8 tubes (one per channel) to deliver fluid
User Interface:
2-line x 24 character LCD screen
25 alphanumeric soft keys
Power Supply:
The instrument uses a 24-volt 90 watt external power supply that
is compatible with 100-240 V~; 50-60 Hz.
Dimensions: 14x10.3x11.5 inches (36x26x29.2 cm) (WxDxH),
Weight: 15 lbs. (6.67 kg)
Operating
Conditions:
10° - 40°C (50° - 104°F)
Relative
Humidity:
The instrument should be operated in a noncondensing humid
environment having a maximum relative humidity of 80% at
temperatures up to 31 decreasing linearly to 50% relative
humidity at 40C.

Table of Contents