62 | Preventive Maintenance
BioTek Instruments, Inc.
Verify Performance
It is important to regularly verify dispense performance. A good test of the MicroFlo’s
performance occurs in the lead up to calibrating the cassette. Perform the initial steps
of one of these calibration procedures to verify performance. The results will
determine the need to actually recalibrate the cassette:
Use the
Cassette Calibration Kit described in the PDF on the Operator’s
Manual CD titled:
7171009_(current Rev)_Calibration Kit
Instructions.PDF
.
Use the
Gravimetric Calibration method described in Appendix C of the
Operator’s Manual.