68 | Preventive Maintenance
BioTek Instruments, Inc.
5. Moisten a cloth with the bleach solution or alcohol. Do not soak the cloth.
6. Wipe the plate carrier, priming trough, dispense arm, and the other exposed
surfaces, except the keypad.
7. Wait 20 minutes. Moisten a cloth with DI or distilled water and wipe all
surfaces of the instrument that have been cleaned with the bleach solution or
alcohol.
8. Use a clean, dry cloth to dry all wet surfaces.
9. Soak the drain tube in hot soapy water or the remaining bleach solution for
twenty minutes or more. Rinse well and set aside to dry.
10. Discard the used gloves and cloths using a Biohazard trash bag and an
approved Biohazard container.