sat801+ Pulse Oximeter
Operation
47
1. Starting from the main menu, call:
Configuration Acoust.
Alarms Silence Time
2. Set the alarm tone silence time
between 30 and 120 sec.
3. Confirm this setting with RETURN.
NOTICE!
During the alarm silence time, the
display shows the alarm silence
symbol: .
Fig. 47: Alarm tone silence time
CAUTION! New alarms cannot be detected!
If a new alarm condition occurs during the alarm tone mute duration, the new alarm
will be displayed after completion of the mute duration.
7.6.3 Masimo Signal Processing (Signalprocessing)
Starting from the main menu, call:
Configuration Signalprocessing
Following actual signal processing
parameters can be set:
– Averaging time (Averaging time)
– Sensitivity (Algorithm mode)
– Smart Tone (on/off)
– Artifact filter (on/off)
Fig. 48: Masimo signal processing
7.6.3.1 Averaging time (Averaging time)
The averaging time indicates the duration, with which from several original measured values
individual displays one VALUE is again computed.
The averaging time can be adjusted between 2-4 (FastSat
TM)
and 16 seconds. 8 seconds is the
default setting.
The longer the length of time, in which measured values are collected, the lower the display value
varies.