sat801+ Pulse Oximeter
Operation
45
NOTICE!
An invalid access code is answered
by the error message: "Incorrect
access code! “ Re-enter the
correct code.
Fig. 43: Invalid access code
7.6.2 Acoustical alarms (Acoust. alarms)
Starting from the main menu, call:
Configuration Acoustic. alarms
Following settings can be adjusted:
– Pulse tone volume (Volume
Pulse)
– Alarm tone volume (Volume
Alarm)
– Alarm tone mute time (Silence
Time)
Fig. 44: Acoustical alarm
7.6.2.1 Pulse tone volume (Volume Pulse)
When the pulse signal tone is activated, then in monitoring mode a tone signal will be played for
every detected pulse. The pitch of the tone identifies the current oxygen saturation. In other
words, the higher the pitch of the tone, the greater the measured oxygen saturation (and vice
versa).
1. Starting from the main menu, call:
Configuration Acoust.
Alarms Volume Pulse
2. Adjust the pulse tone volume from
OFF und loud (10).
NOTICE!
Every incremental change emits the
actual volume strength once.
3. Confirm this setting.