sat801+ Pulse Oximeter
Cleaning and Maintenance
67
11 Cleaning and Maintenance
11.1 Cleaning
11.1.1 Cleaning the Unit
ATTENTION! Danger to the Unit!
Do NOT use strong cleaning agents with petroleum base or acetone solution.
Proceed particularly cautiously in the area of the LCD display to avoid scratching the
surface.
Never immerse the unit in or under water i.e. into another liquid.
The unit as well as the applicable Masimo sensors do not sustain autoclave, steam
or gas sterilization.
1. Before cleaning turn the unit OFF and disconnect from the network and the patient.
2. Only clean the unit with a dry or slightly moist cloth. For more severe soiling use a moist
cloth and all-purpose cleaner.
3. Let the unit dry completely before re-use.
11.1.2 Cleaning of the Masimo-Sensors
The re-usable Masimo sensors are to be cleaned as follows:
1. Disconnect from the patient cable and also from the patient.
2. Use a cloth rinsed in 70% isopropyl alcohol to wipe the entire sensor.
3. Let the entire sensor completely air dry long enough before reusing.
NOTE!
The single-use sensors are only intended for one patient and may not be cleaned or re-
used.