After input 0mV, select ok, the user can see the below screen.
After successful calibration lower AD calibration the system will requests the user to input higher
range input to the input channel.
After input 60mV, press ok, the user can see the below screen.
After successful calibration upper AD calibration the system will show the calibration completed
message.