2.3.6 Electric drive system
The bed’s electrical drive system is first-error-secure, flame-resistant (UL94V-0) and
consists of:
• Electricity supply via:
o 230 V mains cable connection (optional). In this case, the 230 V mains
supply is transformed to 24 V by the control unit.
or via
o an “external” power adapter (optional). The power adapter consists of a
voltage transformer and a low-voltage connection cable. The voltage
transformer generates a protective low voltage that is safe for residents and
users. The power adapter supplies a protective low voltage to the control unit
of all drives (motors), via a connection cable.
• The central control unit. All drive motors and the handset are connected to the
central control unit via plug connections which work with protective low voltage.
• Electric motors for the backrest and thigh rest
• Two electric motors for adjusting the height of the mattress base
• A handset with an elastic hook
2.3.7 Special electrical features (optional equipment)
- Electrical setting of Trendelenburg position: An external operating device (handset)
allows medical staff to place the resident in an emergency position whenever necessary.
- Rechargeable battery for unrestricted electric emergency operation with full lifting capacity
for electric adjustments if a mains connection is not available: can be supplied ready
integrated or can be easily retrofitted for temporary use.
- LED reading lamp: Energy-saving, no hazardous heating as with conventional lamps,
resistant to jolts and vibrations; approx. 50,000-hour lifetime of the LED bulb.
- Discreet LED night light under the bed provides orientation for the resident and prevents
falls from occurring at night.