5 Cleaning and disinfection
This bed is not suitable for machine washing or for cleaning in a
decontamination unit. The bed is only suitable for manual cleaning and
disinfection. To extend the bed's service life and preserve its operability,
always follow the instructions given in this chapter.
5.1 GENERAL INFORMATION ON CLEANING AND DISINFECTION
Cleaning is the most important measure and requirement for ensuring successful
chemical disinfection.
When the bed is occupied by the same resident, routine cleaning of the bed once every
14 days, or as required, is generally sufficient. Disinfection of the chassis is only
necessary if it has been visibly contaminated with infectious or potentially infectious
materials (blood, stool, pus etc.) or if the doctor requires this due to the presence of an
infectious disease.
Before a new resident occupies the bed, the bed must first be cleaned and
disinfected by wiping!
Before cleaning or disinfecting:
Unplug the power adapter from the mains socket and store the adapter
so that it does not come into contact with water or cleaning solutions.
Ensure that none of the electrical components show any signs of external
damage. Failure to comply with this may result in the ingress of water or
detergents into the electronic equipment and cause malfunctions or
damage.
Before using the power adapter again, ensure that there is no residual
moisture on the electrical contacts.
The electrical components must not be cleaned with a water jet, a high
pressure cleaner or other similar devices! Clean only with a moist cloth!
If you suspect that water or any other form of moisture has penetrated the
electrical components, unplug the power adapter immediately or, if
already unplugged, do not plug it back into the mains socket. Label the
bed clearly as “Out of Order” and take the bed out of service. Report this
immediately to the operator responsible.
Failure to follow these safety instructions could result in considerable
damage to the equipment and lead to subsequent malfunctions!