4.4.2 Handset
The electrical bed functions can be activated by the resident or the user by using the
handset. For safety reasons, the handset is equipped with a locking function.
Depending on the clinical condition of the patient, the user must lock handset
adjustments when deemed necessary by the supervising doctor (chapter 4.4.4).
- The electric motors operate only as long as
the corresponding buttons are held pressed.
- Adjustments are possible in both directions.
- An elastic hook allows the handset to be
hung at practically any position on the bed.
- The coiled cable provides ample flexibility
and freedom of movement.
- The handset can
be wiped clean.
- The following basic rule applies to the
buttons:
Only one button can be pressed at a time, otherwise all adjustments stop
(emergency off safety function).