9.3 ELECTRICAL DATA
Control unit for external power adapter:
Type
LINAK CB06 OpenBus
®
Input voltage
20-34 V DC
Max. current input
DC 8 A
Internal device fuse
2-fold electronic, automatic-resetting overload protection;
Duty cycle
Duty cycle: 2 min ON / 18 min OFF
Protection category
IP X6
Classification
Protection Class II, Type B, MPG classification Class I,
not for use in explosive atmospheres
Control unit with internal transformer
Type
LINAK CB06 OpenBus
®
Input voltage
AC 230 V, ± 10 %, 50/60 Hz
Max. current input
AC 2 A
Output voltage
DC 24 V
Internal device fuse
2-fold electronic, automatic-resetting overload protection;
Duty cycle
Intermittent duty 2min ON / 18min OFF
Protection category
IP X6
Classification
Protection Class II, Type B, MPG classification Class I,
not for use in explosive atmospheres
External power adapter
Type
LINAK SMPS 20
Input voltage
AC 230 V, ± 10 %, 50/60 Hz
Mains plug of power
adapter
2-pole Europlug (EN50075)
Current input
AC max. 2.5 A
Output voltage
DC 30 V
Output current,
electronically limited,
Max. 10 A
Activated via remote signal
Duty cycle
Intermittent duty: 2 min ON / 18 min OFF
Protection category
IP X4
Classification
Protection class II, type B,
not for use in explosive atmospheres
Output cable
Approx. 2.5 m, with special plug, 4-pole