7 Troubleshooting
The following table provides a guide for rectifying common malfunctions. Should a
malfunction occur that is not included in the table, inform your operator.
Problem
Possible causes
Solution
Handset/
drive system
not functioning
(bed is connected to
the mains power
supply)
The power adapter is not
plugged in or not plugged in
correctly
No power supply to mains
socket
Functions are locked on handset
Handset, power adapter or
control unit is defective
Serious error/motor fault has
occurred
Plug in the power adapter;
green LED on the power adapter
must light up
Check the mains socket and fuse
box
Unlock the functions (see chapter
4.4.4)
Inform your operator to arrange
for the necessary repairs
Perform RESET; see chapter
6.5.5 / replace motor
Green LED on
external power
adapter* does not
light up and drive
system does not work
The power adapter is not
inserted properly
No power supply to mains
socket
The power adapter is faulty/not
inserted properly
Plug in the power adapter
Check the mains socket
Replace the power adapter with a
new one; inform your operator to
arrange for the necessary repairs
If an external power
adapter* is used:
Drives stop suddenly
after lengthy period
of adjustment
Thermal switch in power adapter*
was triggered by overload
Do not make continuous bed
adjustments for more than 2 minutes!
After 2 minutes of continuous
operation, a break of at least 18
minutes must be observed
To reset the power adapter after an
overload:
Disconnect the device from the power
supply and let it cool down for at least
30 minutes. Then reconnect the
device to the power supply. If the
device still does not function: Device
is faulty replace the device
Yellow LED does not
light up when
pressing the button
on the handset/lights
up green constantly
Functions are locked on handset
Handset faulty
The cable from the power
adapter (24 volt) is not plugged in
properly
Replace handset
Check plug connections
Unlock the adjustment functions
(see chapter 4.4.4)
Handset not
functioning,
adjustments are not
locked
Handset faulty
The control unit has detected
a fault and for safety reasons
has locked the adjustment
functions
Replace handset.
Perform RESET; see chapter 6.5.5;
If it occurs again: Have drive
system checked. Inform your
operator to arrange for the
necessary repairs