Instruction Manual for Lenus Care Bed
Page 82
Problem
Possible causes
Solution
Operation is not
possible despite
functioning power
supply
Control unit has shut down due
to overheating
The control unit has detected a
fault and for safety reasons has
locked the adjustment functions
Control unit defective
Observe max. duty cycle:
After 2 min ON, 18 min OFF;
replace the control unit.
Perform reset, see chapter
6.5.5.
Replace control unit.
Inform your operator to
arrange for the necessary
repairs
Drive runs for a brief
time only, then stops
Drive overloaded
Structural obstructions in the
way of bed adjustment
Remove the overload in the
bed, retest
Remove obstructions; move
bed away from obstructions
(e.g. windowsills, sloping roofs)
Control unit partly
not functioning;
One or more motors are not
connected properly/ electrical
plug connections have come
loose
There is a serious problem with
the control unit.
For safety reasons, all functions
are locked.
Check electrical connection of
all motors/plug connections;
Perform RESET (see chapter
6.5.5); if it occurs again: Have
drive system checked.
Inform your operator to
arrange for the necessary
repairs
Height adjustments
and tilting not
possible or only in
one direction; signal
tone sounds during
adjustment
Control unit has “forgotten” the
drive positions
Perform initialisation; see
chapter 6.5.4.
Pulsating beeping
sound when button
pressed when bed
connected to power
supply
Motor fault
Handset fault
Replace the motor
Replace handset with a new
one