6.5.6 Replacing the motors
Proceed as follows:
Unplug the power adapter from the mains socket.
Remove the faulty motor.
Attention: Raised rests could drop down when the motor is being removed.
Hold the rests securely when you remove the motor.
Unplug the plug-in cable connection on the motor.
Fit the new motor in exactly the same way. Attention: Secure the motor in place
using the motor bolt.
After this, check the electrical bed adjustment functions (see chapter 6.2).
6.5.7 Decommissioning
If the bed is not used for an extended period, please follow the instructions below for
taking the bed out of service safely and ensuring ideal conditions for its re-use:
1. Clean and disinfect the bed see chapter 5) and cover it to protect it against new
contamination.
2. Adjust the mattress base to a flat home position at its lowest level.
3. Lock the electric adjustment functions to prevent them from being activated
accidentally or by unauthorised persons (see chapter 4.4.4).
4. Engage the brakes on the bed.
5. Pay attention to the ambient conditions required for storage (see chapter 9.4).
Please note: The following procedure applies only to the replacement of the backrest
motor and thigh rest motor. Replacing the lift motors is more complex and requires
expert knowledge.
In this case, contact Burmeier’s customer service. We will replace the motors and
dispose of the old motors properly (see chapter 6.4 for the address).