Instruction Manual for Lenus Care Bed
Page 36
• The motor is now disconnected
from the backrest.
• Put the motor down on the bed
frame.
• After the second person has left
the danger zone, the first person
(with the help of the second
person) lowers the backrest
carefully.
Hold the backrest firmly when lowering it, as it could otherwise fall
unchecked!
• Now the motor on the lifting bar is no longer connected to the motor connector
mount.
• The lifting bar remains in the CPR release position.
Restoring the bed to its original state following CPR release of the backrest
• Raise the backrest by hand.
• Swing the lifting bar up again, use the bolt to secure it in place in the motor connector
mount and fold the curved clip back over.