Inspection Report following an Inspection of Electromedical Equipment
according to D
IN EN 62353 (VDE 0751-1): 2015-10 – Page 2 of 3
III. Functional test
What? How?
Functional check of the electrical components
End of travel cut-out of the motors
Handset, control units, external power
adapter
Test according to instruction
manual: Locking functions; button
function; no “rattling” when shaken
No abnormal noise development
(rattling, uneven running etc.)
Installation of cable harness and fixing
of plugs and strain relief
Secure attachment, firm fixing acc.
to operating instructions
Functional check of the mechanical components
Braking, secure engaging of
brakes
Secure engagement, unlocking acc.
to operating instructions
Accessories (e.g. patient lifting pole,
grab handle)
Secure attachment, without
damage, suitability for bed
Inspection passed; test approval sticker applied:
No safety-related or functional defects were detected
No direct risk; the defects detected can be rectified quickly
Inspection was not passed; no test approval sticker applied:
Appliance must be taken out of circulation until the defects have been rectified!
Device does not meet requirements – modification / replacement of components / decommissioning is
recommended!
Documents that form part of this inspection report:
Enclosure, page 3/3: Dimensional check of safety sides in compliance with statutory regulations