EasyManua.ls Logo

burmeier Lenus - Page 71

burmeier Lenus
100 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instruction Manual for Lenus Care Bed
Page 71
Inspection Report following an Inspection of Electromedical Equipment
according to D
IN EN 62353 (VDE 0751-1): 2015-10 Page 2 of 3
III. Functional test
What? How?
OK
Not
OK
Description
of defect
Functional check of the electrical components
End of travel cut-out of the motors
Automatic cut-out
Handset, control units, external power
adapter
Test according to instruction
manual: Locking functions; button
function; no “rattling” when shaken
Motors
No abnormal noise development
(rattling, uneven running etc.)
Installation of cable harness and fixing
of plugs and strain relief
Secure attachment, firm fixing acc.
to operating instructions
Functional check of the mechanical components
Joints and pivots;
Smooth operation
Castors
Braking, secure engaging of
brakes
Safety sides
Secure engagement, unlocking acc.
to operating instructions
Lower leg rest
Engages properly
Accessories (e.g. patient lifting pole,
grab handle)
Secure attachment, without
damage, suitability for bed
Inspection result:
Inspection passed; test approval sticker applied:
No safety-related or functional defects were detected
No direct risk; the defects detected can be rectified quickly
Inspection was not passed; no test approval sticker applied:
Appliance must be taken out of circulation until the defects have been rectified!
Device does not meet requirements modification / replacement of components / decommissioning is
recommended!
Next inspection date:
Documents that form part of this inspection report:
Enclosure, page 3/3: Dimensional check of safety sides in compliance with statutory regulations
Inspected on:
Inspected by:
Signature:
Evaluated on:
Operator/Expert:
Signature:

Table of Contents

Related product manuals