Instruction Manual for Lenus Care Bed
Page 78
- There must be no pauses longer than 6 seconds between the steps without
pressing a key. The system will otherwise revert to the normal operating mode
without completing the initialisation.
How should this be done?
Press and hold both of the marked keys simultaneously
(a rapid intermittent signal sounds) until after
approximately 5 seconds a slow intermittent signal
sounds (=RESET/ manual mode)
When the lift motors are at their end position
(top/bottom), the initialisation is complete.
Otherwise, follow step 2 below.
Raise the mattress base
height with the “Height UP” button to
the highest horizontal position until
both motors switch off automatically at the highest
position, and keep the button pressed for a further 2
seconds. When the button is released, the intermittent
sound signal stops.
------- End of initialisation ------------
If no adjustments are possible and intermittent sound signals sound instead, this
means that the control unit has detected a fault and is locked.
➢ In this case only, carry out the following:
“RESET the control unit” (see chapter 6.5.5)