Cont1= (Continuous Output on Serial Interface)
Press the TARE key to show the current setting. If the setting displayed is acceptable,
press the TARE key again to save it.
Otherwise, use the Fn/ key to toggle the setting and press the TARE key to save it.
Cont1=YES
Cont1=No
Continuous Output on Serial Interface
No Continuous Output on Serial Interface
If Cont1=YES (Continuous Output) is selected, an additional prompt, tYPE= will be
displayed.
If Cont1=no (No Continuous Output) is selected, proceed to the Weight-On- Demand
section.
tyPE= (Continuous Output Format)
Press the TARE key to show the current setting. If the setting displayed is acceptable,
press the TARE key again to save it.
Otherwise, use the Fn/ key to toggle the setting and press the TARE key to save it.
0 = Continuous Output uses SMA format
1 = Continuous Output uses Cardinal Scoreboard format
SMA Continuous Output Format
If SMA is selected, data will be transmitted in the following format:
<lf><s><r><n><m><f><xxxxxx.xxx><uuu><cr>
Where:
= Flags
Z= center of Zero, O = Overcap, E = zero Error,
e = weight not currently being displayed
G = Gross, T = Tare, N = Net
M = Motion, " "(blank) = no motion
Ten characters including a decimal point (if any)