8400-M022-O1 Rev K Model 190 Indicator
15
Settings
IMPORTANT! Calibration and Configuration parameters are not stored in the
non-volatile memory until SETUP is exited. If power is lost while in Setup, any
changes made will be lost and the 190 will revert to the previous configuration.
SETUP
USA= (Domestic or International)
With SEtUP displayed, press the TARE key. The display will change to USA=. Press the
TARE key to show the current setting. If the setting displayed is acceptable, press the
TARE key again to save it.
Otherwise, use the Fn/ key to toggle the setting and press the TARE key to save it.
Yes
(Domestic)
no
(International)
trL=
(4% Zero Range) = no
trL=
(4% Zero Range) = yes
CAP
(Capacity) = + 4% to OC
CAP
(Capacity) = + 9 grads to OC
Date Format = MM/DD/YY
Date Format = DD/MM/YY
LFt= (Legal-For-Trade)
Press the TARE key to show the current setting. If the setting displayed is acceptable,
press the TARE key again to save it.
Otherwise, use the Fn/ key to toggle the setting and press the TARE key to save it.
Yes
= Interval Settings (
Int=
)
allowed are: 1, 2, 5, 10, 20, 50
no
= Interval Setting (
Int=
) is
selectable from 1 to 99.
When both
LFt=yes
and
USA=yes
, the followings results occur:
Scale must have between 100 and 10,000 divisions
Inhibit serial data during input
trA=
(Zero Tracking Range) = 0.5 or 0 to 3
trL=
(4% Zero Range) = no
CAP
(Capacity) = + 4% to OC
When
LFt=yes
and
USA=no
, the followings results occur:
UnS=
(Motion Range) = 1
trL=
(4% Zero Range) = yes
CAP
(Capacity) = + 9 grads to OC