8400-M022-O1 Rev K Model 190 Indicator
Unit1= (Weighing Unit 1)
Press the TARE key to show the current setting. If the setting displayed is acceptable,
press the TARE key again to save it.
Otherwise, use the Fn/ key to toggle to a new setting and press the TARE key to save
it. Allowable settings are:
int= (Interval Setting)
Press the TARE key to show the current setting.
If LFt=YES, use the Fn/ key to toggle to a new setting and press the TARE key to
save it. Allowable settings are: 1, 2, 5, 10, 20, or 50.
If LFt=no, use the Fn/ and UNITS/ keys to enter a new setting and press the
TARE key to save it. Allowable settings are 1 through 99.
When the setting displayed is acceptable, press the TARE key again to save it.
dPP= (Decimal Point Setting)
Press the TARE key to show the current setting. If the setting displayed is acceptable,
press the TARE key again to save it.
Otherwise, use the Fn/ key to toggle to a new setting and press the TARE key to save
it. Allowable settings are:
CAP= (Capacity)
Press the TARE key to show the current setting. If the setting displayed is acceptable,
press the TARE key again to save it.
Otherwise, use the Fn/ and UNITS/ keys to enter a new setting and press the
TARE key to save it. Allowable capacity settings are 1 through 999,999.
Unit2= (Weighing Unit 2)
Press the TARE key to show the current setting. If the setting displayed is acceptable,
press the TARE key again to save it.
Otherwise, use the Fn/ key to toggle to a new setting and press the TARE key to save
it. Allowable settings are:
NOTE: The selection for Unit2 cannot be the same as Unit1. Also, dependent
upon the selection for Unit1, the interval and decimal point settings, not all unit
combinations are available.