8400-M022-O1 Rev K Model 190 Indicator
Unit3= (Weighing Unit 3)
Press the TARE key to show the current setting. If the setting displayed is acceptable,
press the TARE key again to save it.
Otherwise, use the Fn/ key to toggle to a new setting and press the TARE key to save
it. Allowable settings are:
NOTE: The selection for Unit3 cannot be the same as Unit1 or Unit2. Also,
the prompt for Unit3, will not appear if Unit2 is 0.
trA= (Zero Tracking Range)
Press the TARE key to show the current setting assigned to the Automatic Zero
Tracking Range. This is the value in scale divisions that will be automatically zeroed off. If
the setting displayed is acceptable, press the TARE key again to save it.
Otherwise, use the Fn/ key to toggle to a new setting and press the TARE key to save
it. Allowable values are 0 (disables Zero Tracking), 0.5, or 1 through 9.
trL= (4% Zero Range)
Press the TARE key to show the current setting. If the setting displayed is acceptable,
press the TARE key again to save it.
Otherwise, use the Fn/ key to toggle the setting and press the TARE key to save it.
PUO= (Power-Up Zero Feature)
Press the TARE key to show the current setting. If the setting displayed is acceptable,
press the TARE key again to save it.
Otherwise, use the Fn/ key to toggle the setting and press the TARE key to save it.
Automatic Re-Zero on Power-Up
td = (12 or 24 Time Format)
Press the TARE key to show the current setting. If the setting displayed is acceptable,
press the TARE key again to save it.
Otherwise, use the Fn/ key to toggle the setting and press the TARE key to save it.
NOTE: In the 24-hour format, 12 is added to all times after noon, i.e., 3 PM would be 1500.
12-hour clock (3PM displays 3:00)
24-hour clock (3PM displays 15:00)