8400-M022-O1 Rev K Model 190 Indicator
40
Display High Resolution Weight
IMPORTANT! The Hires (Hires?) mode requires a load of 10% of Capacity on
the scale.
Hi res
High Resolution Weight (from SETUP Prompt)
1. If displaying High Resolution Weight is desired, with the SETUP prompt displayed,
press UNITS/ key until the display shows the Hires prompt.
2. Place a calibrated test weight on the scale and press the TARE key.
3. The display will change to show the amount of the test weight in "high resolution" mode
(1/10 division), an h will be displayed to the right of the annunciators and the
annunciators will alternately flash off and on.
4. Press the Fn/ key to increase the span by 0.1 division OR press the UNITS/ key to
decrease the span by 0.1 division.
5. Press the TARE key to exit the Display High Resolution Weight and advance to
LoCoUt? prompt.
Hires?
High Resolution Weight (after TARE pressed at FSPAN prompt)
1. With Hires? displayed, press the TARE key.
2. The display will change to no.
3. Place a calibrated test weight on the scale.
4. Press the Fn/ key to toggle to yes and press the TARE key.
5. The display will change to show the amount of the test weight in "high resolution" mode
(1/10 division), an h will be displayed to the right of the annunciators and the
annunciators will alternately flash off and on.
6. Press the Fn/ key to increase the span by 0.1 division OR press the UNITS/ key to
decrease the span by 0.1 division.
7. Press the TARE key to exit the Display High Resolution Weight and advance to the
LoCoUt? prompt.