61System Specications
System Specifications
Location Requirements
• Designated area with a sturdy, non-ammable workbench
• The BLADE without racks weighs 130 lbs (59 kg)
• Is free from vibration of large equipment
• Minimum counter space: 35 in (88.9 cm) W x 25 in (63.5 cm) D x 29.5 in (75 cm) H
• BLADE Dimensions: 19 in (48.3 cm) W x 24 in (61 cm) D x 29.5 in (75 cm) H
• The above provides 8” (20 cm) of open space on each side and 1” (2.6 cm) open space behind the
instrument for proper ventilation.
• System exhaust connection exits top of unit. Shelves or cabinetry above unit is not recommended.
• Allows for easy access to the fume cabinet and room for the door to extend fully at 16.5” (42 cm)
• Provides adequate space for sample handling
• Provides clear visual access to the display of the system
• Adequate exhaust that pulls a minimum of 125 CFM at point of connection
• DO NOT install in a fume hood
• DO NOT place items on top of the instrument
• DO NOT position unit such that it is difcult to access the power cord receptacle. The power cord receptacle
must be readily accessible for mains disconnection if needed.
Electrical Requirements
• Region Specic
• 120 VAC, 50/60 Hz, 15 A
• 200 VAC, 50/60 Hz, 10 A
• 240 VAC, 50/60 Hz, 10 A
• Fuse Replacements
• 50/60 Hz, 120 VAC: 250 V, 15 A 6.3 mm x 32 mm, fast acting (F15AH250V)
• 50/60 Hz, 200 & 240 VAC: 250 V, 10 A 6.3 mm x 32 mm, fast acting (F10AH250V)
• Line voltage must not vary more than 10% of its specied level
• Dedicated, grounded outlet no more than 8 feet (2.5 M) from unit
• Instrument is supplied with a region specic power cord or an adequately rated power cord may be used
Environmental Requirements
• Indoor use only
• Altitude up to 2000 m
• A Temperature range of 41 °F (5 °C) to 104 °F (40 °C)
• Relative humidity range of 10% to 85%, non-condensing
• Overvoltage Category II
• Pollution degree 2