[DB600] General instruction manual
- 140 -
Storage condition
-20 to 60°C
* 10 to 30°C for long-term storage
5 to 95% RH (no condensation)
0 m/s
2
(under factory packing condition)
Safety and EMC standards (UL/cUL conformity
pending)
EN61010-1 (CE marking)
EN61010-2-030 (CE marking)
UL61010-1 (UL conformity pending)
UL61010-2-030 (UL conformity pending)
CAN/CSA C22.2 No.61010-1 (c-UL
conformity pending)
Overvoltage category: II
Transient overvoltage: 2500V
Pollution degree: 2
Short-time temporary overvoltage: 1440V
Long-time temporary overvoltage: 490V
EN61326-1 ClassA
* Indication or output value varies by the
amount equivalent to ±10% of FS or
±2mV, whichever is larger, during testing.
Event output (option)
Mechanical relay
DB630
:
2 points maximum
DB650
:
4 points maximum
DB670
:
4 points maximum
Open collector 5 points maximum
Output capacity (mechanical relay output):
Resistive load 240V AC 3A
30V DC 3A
Inductive load 240V AC 1.5A
30V DC 1.5A
Minimum load 5V DC 10mA
Output capacity (open collector output):
Heater disconnection detection (option)
Measure heater current using an external
current transformer (CT) to detect
disconnection.
* Designated CT: produced by U.R.D. Co.,
LTD.
CTL-6-S-H: Hole diameter φ5.8,800 turn
0.0 to 65.0mA AC (50/60Hz)
Heater current
measuring range
Transmission signal output (option)
4 to 20mA DC (load resistance 400Ohm
or less)
0 to 1V DC (load resistance 50kOhm
or more)
0 to 10V DC (load resistance 50kOhm
or more)
Non-isolated between the adjustment
output 1 and 2 (ON-OFF pulse output type
excluded)
Remote signal input (option)
4 to 20mA DC (input resistance approx.
50Ohm)
0 to 1V DC (input resistance approx.
500kOhm)
0 to 10V DC (input resistance approx.
100kOhm)
DC current ±30mA or less
±1.5V DC or less
DC voltage ±20V DC or less
External signal input (option)
DB630 5 points maximum, COM shared
DB650 7 points maximum, COM shared
DB670 7 points maximum, COM shared
External contact
capacity
Constant value operation RUN/READY
switch, AUTO/MAN switch, preset manual,
timer 1, timer 2, alarm event reset,
execution No. selection, program/constant
value operation switch, program operation
RUN/STOP switch, program operation
ADVANCE, program operation RESET,
program pattern selection
Communication interface (option)
Type DB630 RS-485
DB650 RS-422A, RS-485
DB670 RS-422A, RS-485
Protocol MODBUS-RTU, MODBUS-ASCII, Private
(used for digital transmission/digital remote
input)
Function Host communication/digital
transmission/digital remote input