Multi-Channel Sync Hipot Tester 19020/19020-4/19021/19021-4/19022/19022-4
User’s Manual
*SRE?
It reads the value of service request enable register. The output format is <decimal
data> within 0~255.
*STB?
It reads the value of status bit register. The output format is <decimal data> within
0~255.
● SCPI Command
:MEMory:DELete[:NAME] <name>
It deletes the parameter data specified by <name> in main memory. <name> is a string
with maximum 13 characters. .
Example: Input the command “MEM:DEL 123”.
Description: It deletes the parameter data named 123 in main memory.
:MEMory:DELete:LOCation <register number>
It deletes the parameter data specified by <register number> in main memory. <register
number> is an integer with the range between 0~30 where 0 means to clear the memory.
Example: Input the command “MEM:DEL:LOC 1”.
Description: It deletes the parameter data number 1 in main memory.
:MEMory:STATe:DEFine <name>, <register number>
It sets a name for memory specified by <register number>. <register number> is an
integer with the range between 1~30.
Example: Input the command “MEM:STAT:DEF TEST,1”.
Description: It defines the parameter data named TEST of the 1
st
memory set in
main memory.
:MEMory:STATe:DEFine? <name>
It queries the memory’s <register number> specified by <name>.
Example: Input the command “MEM:STAT:DEF? TEST”.
The tester returns “1”.
Description: The returned value “1” indicates the parameter data named TEST is
located at the 1
st
set.
:MEMory:FREE:STATe?
It queries the unused capacity in main memory.
Example: Input the command “MEM:FREE:STAT?”.
The tester returns “27”.
Description: The returned value “27” indicates the remaining data parameters for
set.
:MEMory:NSTates?
It queries the capacity of main memory. The value returned is the parameter maximum
plus 1 for *SAV and *RCL commands.
Example: Input the command “MEM:NST?”.
The tester returns “31”.