Multi-Channel Sync Hipot Tester 19020/19020-4/19021/19021-4/19022/19022-4
User’s Manual
[:SOURce]:SAFety:STEP<n>:OSC:CHANnel[:CLOSe]?
It queries the high voltage output channel status of selected STEP.
Example: Input the command SAF:STEP4:OSC:CHAN?
The tester returns (@001,002,201:210).
Description: The returned result means the high voltage output channels CH1 &
CH2 of Address 0 host in STEP 4 are enabled and the high voltage
output channels CH1~CH10 of Address 2 host are enabled as well.
[:SOURce]:SAFety:STEP<n>:OSC:CHANnel:DEFault:ON
It sets the high voltage output channel using the options in DEF. CHANNELS under
TEST CONTROL for selected STEP.
Example: Input the command SAF:STEP4:OSC:CHAN:DEF:ON.
Description: It means the high voltage output channel is set using the options in
DEF. CHANNELS under TEST CONTROL for STEP 4.
[:SOURce]:SAFety:STEP<n>:OSC:CHANnel:DEFault:STATe?
It queries if the high voltage output channel uses the options in DEF. CHANNELS under
TEST CONTROL for selected STEP.
Example: Input the command SAF:STEP4:OSC:CHAN:DEF:STAT?
The tester returns 1.
Description: The returned 1 means the high voltage output channel are set using
the options in DEF. CHANNELS under TEST CONTROL for STEP 4.
[:SOURce]:SAFety:STEP<n>:OSC:CHANnel:PSC P | S | PS, S | C | SC
It sets the scan channel status for selected STEP.
Example: Input the command SAF:STEP1:OSC:CHAN:PSC P,S
Description: It means to set the scan channel of STEP 1 to PS.
[:SOURce]:SAFety:STEP<n>:OSC:CHANnel:PSC?
It queries the scan channel status for selected STEP.
Example: Input the command SAF:STEP1:OSC:CHAN:PSC?
The tester returns P,S.
Description: It means the scan channel of STEP 1 is set to PS.
[:SOURce]:SAFety:STEP<n>:OSC:CORRection[:CAPacitance]:SAMPle <range>,
<numeric value>
It sets the measurement range and standard capacitance for selected STEP. This
command requires two parameters. The first one is to select measurement range from
1~3 where 1 is the smallest capacitance range. The second parameter is to set standard
capacitance in the unit farad. The valid standard capacitance range is 0.001E-09 ~
500E-09.
Example: Input the command SAF:STEP4:OSC:CORR:SAMP 2,5e-9.
Description: It means OSC is set to use the 2
nd
range for host STEP 4 and the
standard capacitance is 5nF.
[:SOURce]:SAFety:STEP<n>:OSC:CORRection[:CAPacitance]:SAMPle?
It queries the measurement range and standard capacitance for selected STEP. It will
return two values separated by comma (,). The first value is measurement range while