GPIB/RS232 Interface (IEEE-488.2)
the second parameter is standard capacitance in the unit farad.
Example: Input the command SAF:STEP4:OSC:CORR:SAMP?
The tester returns 2,3.000000E-09.
Description: The returned 2 means host STEP 4 use the 2
nd
range for OSC and
the standard capacitance is 3nF.
[:SOURce]:SAFety:STEP<n>:OSC:LIMit[:OPEN] <numeric value>
It sets the percentage of open check when doing OSC for selected STEP. The unit is
100%.
Range: 0.1~1.0
Example: Input the command SAF:STEP4:OSC:LIM 0.3.
Description: It means the percentage of open check when doing OSC for the host
STEP 4 is 30%.
[:SOURce]:SAFety:STEP<n>:OSC:LIMit[:OPEN]?
It queries the percentage of open check when doing OSC for selected STEP.
Example: Input the command SAF:STEP4:OSC:LIM?
The tester returns 3.000000E-01.
Description: The returned 3.000000E-01 means the percentage of open check
when doing OSC for the host STEP 4 is 30%.
[:SOURce]:SAFety:STEP<n>:OSC:LIMit: SHORt <numeric value>
It sets the percentage of short check when doing OSC for selected STEP. The unit is
100%.
Range: 0 or 1~5, 0 is OFF.
Example: Input the command SAF:STEP4:OSC:LIM:SHOR 3.
Description: It means the percentage of short check when doing OSC for the host
STEP 4 is 300%.
[:SOURce]:SAFety:STEP<n>:OSC:LIMit: SHORt?
It queries the percentage of short check when doing OSC for selected STEP.
Example: Input the command SAF:STEP4:OSC:LIM:SHOR?
The tester returns 3.000000E+00.
Description: The returned 3.000000E+00 mean the percentage of short check
when doing OSC for the host STEP 4 is 300%.
[:SOURce]:SAFety:STEP<n>:PAuse[:MESSage] <string data >
It sets the message prompt string for PAUSE mode.
Example: Input the command SAF:STEP5:PA “WAIT”.
Description: It means the message string is WAIT for the host STEPS 5.
[:SOURce]:SAFety:STEP<n>:PAuse[:MESSage]?
It queries the message prompt string set.
Example: Input the command SAF:STEP5:PA.
The tester returns “WAIT”.
Description: The returned “WAIT” means the message is “WAIT” for the host