Multi-Channel Sync Hipot Tester 19020/19020-4/19021/19021-4/19022/19022-4
User’s Manual
:SYSTem:TCONtrol:AGC[:SOFTware]?
It queries if the software AGC function is enabled.
Example: Input the command SYST:TCON:AGC?
The tester returns 1.
Description: The returned 1 means the software AGC function is enabled.
:SYSTem:TCONtrol:CHANnel[:DEFault][:CLOSe] <channel_list>
It sets the default of high voltage channel to close.
Example: Input the command SYST:TCON:CHAN (@001:003).
Description: It means the default of high voltage channel is set to close for
CH1~CH3 of host Address 0.
:SYSTem:TCONtrol:CHANnel[:DEFault][:CLOSe]?
It queries the default of high voltage channel to close.
Example: Input the command SYST:TCON:CHAN?
The tester returns (@001:003).
Description: It means the default of host high voltage channel is set to close for
CH1~CH3 Address 0.
:SYSTem:TCONtrol:DISCharge:VMINimum ON | OFF | <boolean>
It sets the total discharge function to ON or OFF.
Example: Input the command SYST:TCON:DISC:VMIN ON
Description: It means to turn on the total discharge function.
:SYSTem:TCONtrol:DISCharge:VMINimum?
It queries if the total discharge is turned on.
Example: Input the command SYST:TCON:DISC:VMIN?
The tester returns 1.
Description: It means the total discharge function is turned on.
:SYSTem:TCONtrol:EOT TIMer | TEST
It sets the EOT signal to be End Of Test or End Of Timer.
Example: Input the command SYST:TCON:EOT TIM
Description: It means to set the host in End Of Timer state.
:SYSTem:TCONtrol:EOT?
It queries the meaning of EOT.
Example: Input the command SYST:TCON:EOT?
The tester returns TIMer.
Description: The returned TIMer means the host is in End Of Timer state.
:SYSTem:TCONtrol:FAIL:OPERation STOP | CONTinue
It sets if continue to test the next step when the present test is judged FAIL.
Example: Input the command SYST:TCON:FAIL:OPER CONT.
Description: It means the test continues when the present test is judged FAIL.