GPIB/RS232 Operation Description (IEEE-488.2)
5-47
5.5 Error Messages
Error #
Label
Message Description
+ 0
No error
No error messages in the queue
-102
Syntax error
The command contains an invalid character
-103
Invalid separator
The command contains an invalid separator
-108
Parameter not allowed
Parameter is invalid and cannot be measured
-109
Missing parameter
Test cannot be performed without missing parameter
-112
Program mnemonic too
long
The header contains more than 12 characters
-113
Undefined header
Header is unrecognizable
-114
Header suffix out of range
Value of numeric suffix is out of range
-120
Numeric data error
Numeric parameter is invalid
-140
Character data error
Input character is invalid
-151
Invalid string data
Incomplete string of data (i.e. missing double “ ”
marks)
-158
String data not allowed
Invalid string of data
-170
Expression error
Incomplete parameter data (i.e. missing right
parenthesis)
-200
Execution error
Invalid execute command
-203
Command protected
Invalid command; device does not receive this
command
-221
Settings conflict
Device is busy and cannot receive command
-222
Data out of range
Parameter value is out of range
-223
Too much data
Received string is too long, device can’t execute
-291
Out of memory
Main memory is full, data cannot be stored
-292
Referenced name does
not exist
Referenced name not found in main memory
-293
Referenced name already
exists
Memory already contains a file labeled ‘referenced
name’
-350
Queue overflow
More than 30 messages in error queue
-361
Parity error
Program message contains parity error
-363
Time out error
End character not received in time, device timed out
-365
Input buffer over run
Input buffer is out of range
-400
Queue error
Output buffer is out of range
-410
Query Interrupted
No read command following a query command
-420
Query Unterminated
No data in queue and device requested to read
queue
The Error Log contains messages that apply to the operation of the instrument with an
RS-232 or GPIB (IEEE-488) Interface installed. Refer to the list of these error messages
below. When there is an error in the queue, the ERROR LED in the lower right-hand corner
of the display will be illuminated. To access the ERROR LOG, press the [MENU] function
key at the top left hand corner of the keypad and press the numeric key [7].
Error messages for the instrument are saved in the queue and are accessed by the FIFO
method. The first error message is the first saved. When the messages in the queue reach
31, the message in the last position will display-350, Queue Overflow”.
Reading a message from the Error Log deletes the error message from the queue.