EasyManua.ls Logo

Chroma 19032 - 1 Introduction; Product Overview; Features

Default Icon
138 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Introduction
1-1
1. Introduction
1.1 Product Overview
This instrument is designed for automatic dielectric withstand voltage (Hipot), insulation
resistance, grounding resistance, open/short circuit detection, and dynamic leakage current
testing of electromechanical and electronic equipment.
Withstand voltage output power: AC: 200VA (5kV, 40mA). DC: 72VA (6kV, 12mA).
Insulation resistance measurement range: 0.1M 50G. Test voltage range: 50V 1000V.
Grounding resistance measurement range: 0.1 150m (under 10A the range is extended to
510m.) Output test current range: 1 30A.
Dynamic leakage current measurement range: 0.01mA 50.0mA (rms). Output test voltage
range: 90V 280VAC. (The test rules conform to IEC950, UL544, and UL2601.)
When testing for open/short circuit issues, obtain the standard capacitance value before
testing with high voltage. Ensure the DUT is properly connected before beginning a high
voltage test.
Display settings include: test mode, status, time, current, voltage, resistance value, and
memory number.
The analyzer is equipped with PASS/FAIL logic and provides signal output of testing results
and allows remote control operation. The analyzer includes an RS-232 interface (standard),
GPIB interface (optional), and SCANNING interface (optional), for connection to automated
test systems.
1.2 Features
AC / DC withstand voltage, insulation / grounding resistance, open/short circuit detection,
and dynamic leakage current scan test (option) in one model.
Simultaneous twin port output.
Dynamic leakage current simulation compensation.
Open/short circuit detection.
Reformation DC quick discharge.
Follows UL, CSA, TUV, CDE, and CE test rules.
0.4ms cut off and 0.2sec discharge rates.
Keypad lock and material protection function.
Eight judgment result indication windows.
Charge current low limit detection function.
Allows combining a total of 500 test procedures or 100 sets of memory functions.
GPIB interface (optional).
Dynamic high voltage leakage current automatic scan function (optional).
Full-function front panel calibration.