Calibration Step
7-7
Press [1] [0] [0] [0] [ENTER]
Press [STOP] ; Stop
Change the standard load resistor to 100M.
Press [] key to display.
Display IRR range2 (100M) ; Resistor of IR to 100.0M
Press [STOP] [START] ; Read out the IRR value
; For example 100.0M
Press [1] [0] [0] [ENTER]
Press [STOP] ; Stop
Change the standard load resistor to 10M.
Press [] key to display.
Display IRR range3 (10M) ; Resistor of IR to 10.0M
Press [STOP] [START] ; Read out the IRR value
; For example 10.0M
Press [1] [0] [ENTER]
Press [STOP] ; Stop
Change the standard load resistor to 10M.
Press [] key to display.
Display IRR range4 (10M) ; Resistor of IR to 10M
Press [STOP] [START] ; Read out the IRR value
Press [1] [0] [ENTER] ; For example 10.0M
Press [STOP] ; Stop
7.8 Complete Calibration
Press [EXIT]
[MENU]
MAIN MENU