GPIB/RS232 Operation Description (IEEE-488.2)
5-45
Return message “1”
Description: Return message “1” means Ramp Judg. is set to ON.
[:SOURce]:SAFEty:PRESet:SCREen < boolean > | ON | OFF
This command enables or disables the Screen function.
Example: Input command “SAFE:PRES:SCRE OFF”.
Description: This command turns OFF the analyzer screen.
TRIGger:SOURce:EXTernal:STATe < boolean > | ON | OFF
This command enables or disables the START KEY when the unit is operating
under remote control.
The START KEY will be enabled when the unit is operating under remote control
when the parameter is set to ON (1).
The START KEY will be disabled when the unit is operating under remote control
when the parameter is set to OFF (0).
Example: Input command “TRIG:SOUR:EXT:STAT 0”
Description: This command disables the START KEY when the unit is operating
under remote control.
TRIGger:SOURce:EXTernal:STATe?
This command queries the enabled/disabled status of the START KEY when the
unit is operating under remote control.
Example: Input command “TRIG:SOUR:EXT:STAT?”
Return message “0”
Description: Return message “0” means the START KEY will be disabled when the
unit is operating under remote control.
5.4.3 Start Test by Serial No.
This analyzer will start a test when the remote interface receives a string that matches a DUT
Serial Number. For example: if the DUT Serial No. has been set to “AA*****” (* refers to a
variable character) and “AA00001” or “AA00300” is received from the remote interface, the
test will be started.