Calibration Step
7-1
7. Calibration Step
The analyzer should be warmed up for at least 30 minutes before starting the calibration
procedure in this section. Remove the calibration label from the front panel and press the
calibration switch once. When the calibration has been completed, press the calibration
switch on the front panel again (this switch is equipped with a hardware protection circuit to
avoid calibration data loss).
Voltage Calibration (See 7.2)
ACV 5kV Offset (500V) ;AC Voltage OFFSET point
ACV 5kV Full (4kV) ;AC Voltage FULL point
DCV 6kV Offset (500V) ;DC Voltage OFFSET point
DCV 6kV Full (4kV) ;DC Voltage FULL point
IRV 1kV Offset (500V) ;IR Voltage OFFSET point
IRV 1kV Full (1kV) ;IR Voltage FULL point
Current Calibration (See 7.3)
ACA 3mA Offset (0.12mA) ;AC 2.99mA range OFFSET point
ACA 3mA Full (2.5mA) ;AC 2.99mA range FULL point
ACA 40mA Offset (2.5mA) ;AC 40mA range OFFSET point
ACA 40mA Full (25mA) ;AC 40mA range FULL point
DCA 3mA Offset (0.12mA) ;DC 2.99mA range OFFSET point
DCV 3mA Full (2.5mA) ;DC 2.99mA range FULL point
DCV 12mA offset (2.5mA) ;DC 20mA range OFFSET point
DCV 12mA Full (10mA) ;DC 20mA range FULL point
Grounding Mode Calibration (See 7.4)
GBA 30A Offset (3A) ;GB current OFFSET point
GBA 30A Full (25A) ;GB current FULL point
GBV 8V Offset (0.3V) ;GB voltage OFFSET point
GBV 8V Full (3V) ;GB voltage FULL point
Withstanding Voltage Mode Arcing Calibration (See 7.5)
AC ARC 40mA(5mA) ;AC Arcing Calibration
DC ARC 12mA(5mA) ;DC Arcing Calibration
Insulation Resistance Mode Leakage Current Meter Calibration (See 7.6 & 7.7)
IRR Range 1 (1GΩ) ;IR Resistor range 1 Calibration
IRR Range 2 (100MΩ) ;IR Resistor range 2 Calibration
IRR Range 3 (10MΩ) ;IR Resistor range 3 Calibration
IRR Range 4 (10MΩ) ;IR Resistor range 4 Calibration
7.1 Calibration
Press [3] [ENTER]
Display “password”
Press [7] [9] [3] [1] [ENTER]
Press [DEVICE]