Electrical Safety Analyzer 19032 Users Manual
6-4
8. LDCHI(?) {f ¦*}
Function: Sets High Limit of LDC.
Parameter: 1. “*”: Disabled - no LDC High Limit test is done.
LDC Mode
f = 0.001 1.000 mA
9. DISP(?) {n ¦c}
Function: Displays RMS or PEAK for LC leakage current.
Parameter: 0 / RMS = RMS
1 / PEAK = PEAK
Description: This parameter can be set only for 6000-08 testing.
10. FALL(?) {f ¦*}
Function: Sets the voltage fall time (the period of time for the voltage to decrease back to
0).
Parameter: f: 0 999 second
Description: FALL can only be set for AC/DC/IR testing.
11. IRRNG(?)
{
n ¦c
}
Function: Selects IR mode RANGE.
Parameter: 0 = AUTO
1 = 300nA
2 = 3uA
3 = 30uA
4 = 300uA
5 = 3mA
6 = 10mA