Electrical Safety Analyzer 19032 User’s Manual
7-2
7.2 Voltage Calibration
7.2.1 ACV Calibration
Connect an ACV high voltage meter to the analyzer or connect a Chroma ATE 9102 Hipot
Calibrator and select ACV MODE [100MΩ].
Display ACV 5kV Offset (100V) ; ACV OFFSET POINT calibration
Press [STOP] [START] ; Read the HV meter value
; Example 0.092kV
(Using the keypad, enter the reading from the voltmeter)
Press [0] [.] [0] [9] [2] [ENTER]
Press [STOP] ; Stop ACV OFFSET POINT calibration
Press [] key
Display ACV 5kV Full (4kV) ; ACV FULL POINT calibration
Press [STOP] [START] ; Read the HV meter value
; Example 4.052kV
(Using the keypad, enter the reading from the voltmeter)
Press [4] [.] [0] [5] [2] [ENTER]
Press [STOP] ; Stop ACV voltage calibration
7.2.2 DCV Calibration
Connect a DCV high voltage meter to the analyzer or connect a Chroma ATE 9102 Hipot
Calibrator and select DCV MODE [1.00GΩ].
Press [] key
Display DCV 6kV Offset (100V) ; DCV OFFSET POINT calibration
Press [STOP] [START] ; Read the HV meter value
; Example 0.092kV
(Using the keypad, enter the reading from the voltmeter)
Press [0] [.] [0] [6] [2] [ENTER]
Press [STOP] ; Stop DCV OFFSET POINT calibration
Press [] key to display
Display DCV 6kV Full (4kV) ; DCV FULL POINT calibration
Press [STOP] [START] ; Read the HV meter value
; Example 4.052kV
(Using the keypad, enter the reading from the voltmeter)
Press [4] [.] [0] [5] [2] [ENTER]
Press [STOP] ; Stop DCV Voltage calibration
7.2.3 IR Voltage Calibration
Connect a DCV High Voltage meter to the analyzer or connect a Chroma ATE 9102 Hipot
Calibrator and select DCV MODE [1.00GΩ].